Using the STATus System
STATus Subsystem
Key Type:
Factory Preset: 0
Range:
Questionable Event Query
:STATus:QUEStionable[:EVENt]?
This query returns the decimal value of the sum of the bits in the
Questionable Event register.
The register requires that the associated PTR or NTR filters be set
NOTE
before a condition register bit can set a bit in the event register.
The data in this register is latched until it is queried. Once queried, the
register is cleared.
Key Type:
Questionable Negative Transition
:STATus:QUEStionable:NTRansition <number>
:STATus:QUEStionable:NTRansition?
This command determines which bits in the Questionable Condition
register will set the corresponding bit in the Questionable Event
register when the condition register bit has a negative transition (1 to
0). The variable <number> is the sum of the decimal values of the bits
that you want to enable.
Key Type:
Factory Preset: 0
Range:
Questionable Positive Transition
:STATus:QUEStionable:PTRansition <number>
:STATus:QUEStionable:PTRansition?
This command determines which bits in the Questionable Condition
register will set the corresponding bit in the Questionable Event
register when the condition register bit has a positive transition (0 to 1).
The variable <number> is the sum of the decimal values of the bits that
you want to enable.
374
There is no equivalent front-panel key.
0 to 32767
There is no equivalent front-panel key.
There is no equivalent front-panel key.
0 to 32767
Chapter 6