Key Path:
BW/Avg
Saved State:
Saved in Instrument State
Factory Preset: Auto (Log-power)
Remote Command:
[:SENSe]:AVERage:TYPE RMS|LOG|SCALar
[:SENSe]:AVERage:TYPE?
[:SENSe]:AVERage:TYPE:AUTO OFF|ON|0|1
[:SENSe]:AVERage:TYPE:AUTO?
Example:
AVER:TYPE:RMS
AVER:TYPE:SCAL
AVER:TYPE:LOG
2.3.5.1 Auto
Chooses the optimum type of averaging for the current instrument measurement settings.
selects Power (RMS) Averaging if
Auto
is set to
and
Man
Average
or Detector is set to Quasi Peak or EMI Average. For other conditions,
Log-Power Average.
Key Path:
BW/Avg
Remote Command:
See
"Avg/VBW Type" on page
Example:
AVER:TYPE:AUTO ON
Chapter 2
or
Auto Couple
Sets Power (RMS) averaging
Sets Voltage averaging
Sets Log-Power (video) averaging
Marker Noise
. It selects Voltage Averaging if
,
Avg/VBW Type
74.
Instrument Functions: A - L
is on,
Band/Intvl Power
Amplitude, Scale Type
Auto
BW/Avg
is on, or
Detector
is set to
Lin
selects
75