Using the STATus System
Common Using the STATus System IEEE Commands
For example, if you want to verify the completion of both calibration
and waiting for trigger set :STAT:OPER:ENAB 33 and monitor any
changes.
Key Type:
There is no equivalent front-panel key.

Query Instrument Options

*OPT?
For ESA and PSA analyzers: This command is not implemented.
However the information is available remotely by using SCPI
commands. See :SYSTem:OPTions? in the SYSTem subsystem.

Power-On Status Clear

*PSC
*PSC?
Sets the state of the power-on status clear flag. This command allows
you to specify if the service request enable register and the event status
enable register should be cleared at power up.
The query returns the contents of the power-on status clear flag.
Key Type:
There is no equivalent front-panel key.