Agilent Technologies PSA Series User And Programming Manual page 237

Spectrum analyzers
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Remote Command:
[:SENSe]:SWEep:EGATe:POLarity NEGative|POSitive
[:SENSe]:SWEep:EGATe:POLarity NEGative|POSitive?
Example:
SWE:EGAT:POL NEG
SWE:EGAT:POL?
4.8.5.3 Delay
Controls the length of time from the time the gate condition is satisfied until the gate is
enabled.
Key Path:
,
Sweep
State Saved:
Saved in instrument state
Factory Preset: 57.7 microsecond
Resolution:
100 ns
0.0 µ s to 100 seconds
Range:
Remote Command:
[:SENSe]:SWEep:EGATe:DELay <time>
[:SENSe]:SWEep:EGATe:DELay?
Example:
SWE:EGAT:DEL 100US
SWE:EGAT:DEL?
4.8.5.4 Length
Controls the length of time that the gate is enabled.
Key Path:
,
Sweep
State Saved:
Saved in instrument state
Factory Preset: 461.6 µ s
10 µ s to 0.5 s
Range:
Remote Command:
[:SENSe]:SWEep:EGATe:LENGth <time>
[:SENSe]:SWEep:EGATe:LENGth?
Example:
SWE:EGAT:LENG 1ms
SWE:EGAT:LENG?
Chapter 4
Gate Setup
Gate Setup
Instrument Functions: P - Z