Instrument Functions: A - L
File
2.6.3.1 Load Now
Loads the currently selected file. Displayed settings include name, type, destination, and
path. While the file is being loaded a popup message is displayed "Loading file." After a
successful load, the text message "xxxxxx file loaded" (where xxxxxx is the file name)
appears in the status line. When traces are loaded they always load in
Traces save in TRC format can be loaded individually or as a group. When a trace is
loaded, the state that existed when that trace was saved is loaded along with the trace.
Also, the loaded trace(s) is/are placed in view mode.
To compare two saved traces from different saves, place traces in view mode
NOTE
before saving them. This prevents the trace from being rewritten based on a
state change from subsequent loads.
Key Path:
,
File
Load
Remote Command:
:MMEMory:LOAD:STATe 1,<'file_name'> loads the specified state file into the current
active state of the instrument.
The MMEM:LOAD:STAT command only saves the state of the Signal Analysis mode. Use *RCL
to load all mode states. *RCL <register#> loads the state from the specified internal register
into the current active state of the instrument. The available register numbers are 0 to 127.
:MMEMory:LOAD:TRACe <label>,<'file_name'>
:MMEMory:LOAD:CORRection ANTenna|CABLe|OTHer|USER,<'file_name'>
:MMEMory:LOAD:LIMit LLINE1|LLINE2,<'file_name'>
Remote Command Notes: For the MMEM:LOAD:TRAC <'file_name'> command,
<'file_name'> must include the following:
Complete path
A file extension of TRC
Use all uppercase letters.
Trace labels are: TRACE1|TRACE2|TRACE3
For the MMEM:LOAD:STAT 1 command:
If the firmware revision of the state being loaded is newer than the
firmware revision of the instrument, no state is recalled and an error is
reported.
If the firmware revision of the state being loaded is the same as the
firmware revision of the instrument, all settings of the state will be
loaded.
If the firmware revision of the state being loaded is older than the
firmware revision of the instrument, the instrument will only load the
older settings of the state.