Agilent Technologies PSA Series User And Programming Manual page 64

Spectrum analyzers
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Instrument Functions: A - L
Auto Couple
2.2.3.1 Auto
Selects the LO phase noise behavior to optimize speed or dynamic range for various
instrument operating conditions.
For PSA, the
rules choose:
Auto
, for span ≥ 10.5 MHz or the Res BW > 200 kHz
• Fast Tuning
• Optimize £(f) for f >50 kHz
• Optimize £(f) for f <50 kHz
Key Path:
Auto Couple
Remote Command:
See "PhNoise Opt" on
Example:
FREQ:SYNT:AUTO ON
2.2.3.2 Optimize £(f) for frequencies < 50 kHz
The LO phase noise is optimized for measuring signals with offsets less than 50 kHz from
the carrier, at the expense of phase noise beyond 50 kHz offset.
Key Path:
Auto Couple
Remote Command:
See "PhNoise Opt" on
Example:
FREQ:SYNT 1
2.2.3.3 Optimize £(f) for frequencies > 50 kHz
Optimizes phase noise for measuring offsets more than 50 kHz from the carrier, especially
those from 70 kHz to 300 kHz. Closer offsets are compromised and the throughput of
measurements (especially remote measurements where the center frequency is changing
rapidly), is reduced. There is a graph of typical phase noise performance in the
Specifications Guide.
Key Path:
Auto Couple
Remote Command:
See "PhNoise Opt" on
Example:
FREQ:SYNT 2
64
, for spans >141.4 kHz, and for Res BWs >9.1 kHz
, for all other spans and Res BWs.
,
PhNoise Opt
page
63.
,
PhNoise Opt
page
63.
,
PhNoise Opt
page
63.
Chapter 2

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4440aE4443aE4445aE4448aE4446a

Table of Contents