Instrument Functions: A - L
AMPLITUDE / Y Scale

2.1.13 Atten Step

Permits the selection of 2 dB or 10 dB step resolution for input attenuation.
Key Path:
AMPLITUDE / Y Scale
Saved State:
Saved in instrument state
Factory Preset: 2 dB
Remote Command:
[:SENSe]:POWer[:RF]:ATTenuation:STEP[:INCRement] <integer> (in dB)
[:SENSe]:POWer[:RF]:ATTenuation:STEP[:INCRement]?
Example:
POW:ATT:STEP 10
Sets the Attenuation to 10 dB. The only valid suffix is dB. If no suffix is
sent, dB is assumed.
If a value >5 is entered, 10 is used.
If a value ≤ 5 is entered, 2 is used

2.1.14 Max Mixer Lvl

Enables you to set the relationship between the highest signal that can be displayed (the
reference level) and the input attenuation. The relationship applies whenever the
is set to
Attenuation
Auto
level minus the max mixer level. For example, as the reference level changes, the input
attunator changes to ensure that a signal at the reference level does not exceed the
setting.
Mixer Lvl
Key Path:
AMPLITUDE / Y Scale
State Saved:
Saved in Instrument State
Factory Preset: –10 dBm
Range:
–50 dBm to –10 dBm (internal mixing)
–50 dBm to +10 dBm (external mixing)
Remote Command:
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer] <power>
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer]?
Example:
POW:MIX:RANG -15 dBm
56
. The relationship is that the attenuation is given by reference
Max
Chapter 2