7
Error Information
Detector Error Messages
Detector Error Messages
These errors are detector specific.
UV lamp: no current
Error ID: 7450
The lamp anode current is missing. The processor continually monitors the
anode current drawn by the lamp during operation. If the anode current
falls below the lower current limit, the error message is generated.
Probable cause
1
2
3
4
5
UV lamp: no voltage
Error ID: 7451
The lamp anode voltage is missing. The processor continually monitors the
anode voltage across the lamp during operation. If the anode voltage falls
below the lower limit, the error message is generated.
94
Lamp disconnected.
Top foam removed while lamp is on.
Defective or non-Agilent lamp.
Defective main board.
Defective power supply.
Suggested actions
Ensure the lamp connector is seated firmly.
Please contact your Agilent service
representative.
Exchange the lamp.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Agilent 1260 Infinity VWD User Manual