Lamp; Max-Light Cartridge Flow Cell - Agilent Technologies InfinityLab LC Series User Manual

Diode array detectors
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

1
Introduction
Optical System

Lamp

The light source for the UV-wavelength range is a long-life UV-lamp with RFID
tag. As a result of plasma discharge in low-pressure deuterium gas, the lamp
emits light over the 190 nm to approximately 800 nm wavelength range.
Figure 5

Max-Light Cartridge Flow Cell

The detector allows easy access to flow cells via a cartridge. A variety of optional
flow cells can be inserted using the same quick, simple mounting system.
Max-Light Cartridge Flow Cells for standard and bio-inert applications are
available. For testing of the detector, a Max-Light Cartridge Test Cell is available.
p/n
G4212-60008
G4212-60007
G4212-60032
G4212-60038
G7117-60020
G4212-60011
The optical principle of the Max-Light Cartridge cell is based on opto-fluidic
waveguides. Nearly 100 % light transmission is achieved by utilizing total internal
reflection in a non-coated silica fiber. Compromising refractive index and thermal
effects are almost completely eliminated, resulting in significantly less baseline
drift.
Agilent InfinityLab LC Series Diode Array Detectors User Manual
UV-Lamp
Description
Max-Light Cartridge Cell (10 mm, V(
Max-Light Cartridge Cell (60 mm, V(
HDR Max-Light Cartridge Cell (3.7 mm, V(
ULD Max-Light Cartridge Cell (10 mm, V(
Max-Light Cartridge Cell LSS (10 mm, V(
Max-Light Cartridge Test Cell
) 1.0 µL)
) 4.0 µL)
) 0.4 µL)
) 0.6 µL)
) 1.0 µL)
15