Exchanging a Lamp

When required
Tools required
Parts required
If you want to use the Agilent DAD lamp instead of the VWD lamp, you have to change the
N O T E
lamp settings in the VWD Configuration to lamp type 2140-0590. This ensures that the DAD
lamp's filament heating is operated like in the DAD. The instrument specifications are
based on the VWD lamp.
If the detector has been in use, the lamp may be hot. If so, wait five minutes for lamp
WA R N I N G
to cool down.
Preparations for this procedure:
Turn the lamp OFF.
1200 Series Variable Wavelength Detector User Manual
If noise or drift exceeds application limits or lamp does not ignite.
Screwdriver POZI 1 PT3
Deuterium lamp G1314-60100
1 Press the release buttons and remove the front cover to
have access to the lamp area.
Maintenance
8
85