7
Error Information

Detector Error Messages

Detector Error Messages

These errors are detector specific.

UV lamp: no current

Error ID: 7450
The lamp anode current is missing. The processor continually monitors the
anode current drawn by the lamp during operation. If the anode current
falls below the lower current limit, the error message is generated.
Probable cause
1
2
3
4
5
104
Lamp disconnected.
Top foam removed while lamp is on.
Defective or non-Agilent lamp.
Defective main board.
Defective power supply.
Suggested actions
Ensure the lamp connector is seated firmly.
Please contact your Agilent service
representative.
Exchange the lamp.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Agilent 1200 Infinity Series VWD - User Manual