7
Error Information
Refractive Index Detector Specific Error Messages
Lamp Voltage too Low
Probable cause
1
2
Lamp Voltage too High
Probable cause
1
2
3
Lamp Current too High
Probable cause
1
2
112
Defective main board.
Defective lamp or optics.
Contaminated flow cell.
Defective main board.
Defective lamp or optics.
Defective main board.
Defective lamp or optics.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Flush the flow cell.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.