Lamp Voltage Too Low; Lamp Voltage Too High; Lamp Current Too High - Agilent Technologies Infinity 1260 User Manual

Refractive index detector
Hide thumbs Also See for Infinity 1260:
Table of Contents

Advertisement

7
Error Information
Refractive Index Detector Specific Error Messages

Lamp Voltage too Low

Probable cause
1
2

Lamp Voltage too High

Probable cause
1
2
3

Lamp Current too High

Probable cause
1
2
112
Defective main board.
Defective lamp or optics.
Contaminated flow cell.
Defective main board.
Defective lamp or optics.
Defective main board.
Defective lamp or optics.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Flush the flow cell.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Agilent 1260 Infinity RID User Manual

Advertisement

Table of Contents
loading

Table of Contents