Maintenance; Cleaning - Siemens MAGNETOM Avanto Operator's Manual

4 ch bi breast / 16 ch ai breast
Hide thumbs Also See for MAGNETOM Avanto:
Table of Contents

Advertisement

Component
Needle blocks, needle
guide
Biopsy device
Frame
4 Ch BI Breast,
16 Ch AI Breast
MAGNETOM Avanto, MAGNETOM Verio | syngo MR E11
Print No. MR-04010G.623.22.02.02
6

Maintenance

6.1

Cleaning

All components have to be cleaned, disinfected, or sterilized
immediately after each examination or biopsy. The type of
preparation depends on the respective component.
Cleaning
Pre-cleaning: spraying
device or water jet
Ultrasound bath
Water bath (soft brush,
if necessary)
damp cloth
 CAUTION
Incorrect cleaning, disinfecting, and sterilizing!
Damage to the coil, device defect
Please adhere to the instructions for cleaning, disinfecting,
and sterilizing the biopsy device and the coil.
Never clean and disinfect the coil in immersion baths or under
running water.
Never sterilize the coil.
Disassemble the components of the biopsy device and process
them individually.
Maintenance
Disinfection
immerse fully in solu-
tion and let it take
effect (at least 5 min)
apply solution and let it
take effect
apply solution and let it
take effect
6
Sterilization
Steam sterilization
Steam sterilization
not permitted
53