Component
Needle blocks, needle
guide
Biopsy device
Frame
4 Ch BI Breast,
16 Ch AI Breast
MAGNETOM Avanto, MAGNETOM Verio | syngo MR E11
Print No. MR-04010G.623.22.02.02
6
Maintenance
6.1
Cleaning
All components have to be cleaned, disinfected, or sterilized
immediately after each examination or biopsy. The type of
preparation depends on the respective component.
Cleaning
Pre-cleaning: spraying
device or water jet
Ultrasound bath
Water bath (soft brush,
if necessary)
damp cloth
CAUTION
Incorrect cleaning, disinfecting, and sterilizing!
Damage to the coil, device defect
◆
Please adhere to the instructions for cleaning, disinfecting,
and sterilizing the biopsy device and the coil.
◆
Never clean and disinfect the coil in immersion baths or under
running water.
◆
Never sterilize the coil.
◆
Disassemble the components of the biopsy device and process
them individually.
Maintenance
Disinfection
immerse fully in solu-
tion and let it take
effect (at least 5 min)
apply solution and let it
take effect
apply solution and let it
take effect
6
Sterilization
Steam sterilization
Steam sterilization
not permitted
53