Removing And Returning The Ion Optics Assembly - PerkinElmer Clarus SQ 8 MS Series Hardware Manual

Hide thumbs Also See for Clarus SQ 8 MS Series:
Table of Contents

Advertisement

Clarus SQ 8 MS Hardware Guide
8000 Grade Micro Mesh (Part No. N9303421)
600 grit aluminum oxide in DI Water with a few drops of methanol to make a
paste
Acetone
Methanol
When cleaning internal components it is important to maintain the quality of the
surface finish. Deep scratches or pits can cause loss of performance. Where no specific
cleaning procedure is provided, you should use fine abrasives to remove dirt from
metal components. Recommended abrasives are:
6000 Grade Micro Mesh (Part No. N9303420).
8000 Grade Micro Mesh (Part No. N9303421).
600 grit aluminum oxide in DI Water with a few drops of methanol to make a paste
After cleaning with abrasives, it is necessary to wash all metal components in suitable
solvents to remove all traces of grease, oil and, if micro-mesh is used, rubber. The
recommended procedure is to swill or sonicate the components in a clean beaker of
methanol for at least ten minutes and subsequently to blot them dry with lint-free
tissue. Recommended solvents are:
After the components are reassembled, they should be blown with oil-free nitrogen to
remove dust particles.

Removing and Returning the Ion Optics Assembly

CAUTION
CAUTION
96
To prevent contamination of the mass spectrometer, always wear clean,
lint-free, powder-free nitrile , nylon, or PVC gloves before touching,
removing or replacing parts. Hold the source by its handle only. Never
touch these parts with ungloved (bare) fingers, as this will introduce
contaminants into the system.

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Clarus SQ 8 MS Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

This manual is also suitable for:

Clarus sq 8 tClarus sq 8 s

Table of Contents

Save PDF