System Administration Menu; Hardware Setup Menu; Prime Fluidics Menu And New Reagents Blank - Thermo Scientific 2210AM Manual

Ammonia analyzer
Table of Contents

Advertisement

Hardware Setup Menu

Allows technician to calibrate several hardware components
of the analyzer.
Calibration results are editable by the technician.
Prime Fluidics Menu
Allows user to prime the fluidics lines of the analyzer.
Diagnostics Menu
Diagnostics provide a direct mechanism to individually test
devices such as pumps, relays, optics and stirrer. Turn
various pumps ON and OFF for testing or during adjustment.
Maximum of two devices can be turned on at the same time.
This function also provides a quick check of the optics
subsystem.
System Administration
Menu
Edit User Menu
Allows technician to edit user names, user passwords and
user permissions, for measurement, calibration and validation.
Allows user to edit his or her own user name and password.
Delete User Menu
Allows technician to delete a user from the users list.
Add User Menu
Allows technician to add a new user to the user's list. Up to six
different users may be added.
Factory Settings Menu
Allows technician to restore factory defaults, update factory
defaults and edit factory settings.
Instrument Details Menu
Allows viewing of instrument details:
Software version

Serial number

Model number

Analysis Label Change Menu
Allows technician to edit the analysis label.
Printing Options Menu
Allows technician to print the system configuration, analysis
parameters and method parameters. Also allows technician
to enable/disable online printing. See Terms Definition Section
for the default analyzer parameters listing printouts.
28 |
Menu Overview
Hardware Setup Menu
Calibrate Thermistor
Allows technician to calibrate the thermistor. Only required if
the factory-installed thermistor has been serviced or replaced.
To calibrate the thermistor, user is required to enter the
reaction block temperature using an external reference
thermometer.
Calibrate Pumps
Allows technician to calibrate the flow rate of the pumps.
Calibrate Optics
Allows technician to calibrate the optics.
Calibrate Analog Outputs
Allows technician to calibrate the two analog outputs.
Edit Calibrated Values
Allows technician to edit the calibrated results for fine-tuning.
Prime Fluidics Menu and
New Reagents Blank
Auto Prime All
This menu runs an automatic priming sequence for all
reagents, standards and sample based on the priming volume
specified by the user.
Prime Input Source (Sample/Reagent/...)
Allows technician to selectively prime from the available list of
devices.
Priming Volume
Allows technician to change the volume used for priming.
Press + to Navigate to NEXT Page
The new reagents blank is the ZERO intercept or offset. A
user entry here will affect the zero calibration value of the
calibration parameters.
Set NEW reagents BLANK
Allows the operator to set the BLANK value for the NEW
reagents. This value is usually supplied with the new batch of
reagents and its value may vary between 0 to -10. Default
value is range dependent. PLEASE NOTE THE (-) SIGN.
Orion 2210AM Ammonia Analyzer