Auto Couple

Couples the following functions: resolution bandwidth, video
bandwidth, sweep coupling (SR/SA), attenuation, sweep time,
center-frequency step, and frequency counter resolution.
Coupled functions are functions that are linked. If one function is
NOTE
changed, the coupled function is changed. During normal operation, the
sweep time, resolution bandwidth, and video bandwidth are coupled to
span, the input attenuation is coupled to the reference level, and the
center frequency step is coupled to the span.
This function is not available when the segmented sweep function is set
NOTE
to On (
Chapter 6
).
Segmented (On)
Front-Panel Key Reference
Auto Couple
6-13