Calibrating for Increased Measurement Accuracy

Frequency Response Error Corrections

Frequency Response Error Corrections
You can remove the frequency response of the test setup for the following measurements:
• reflection measurements
• transmission measurements
• combined reflection and transmission measurements

Response Error Correction for Reflection Measurements

1. Press
.
Preset
2. Select the type of measurement you want to make.
❏ If you want to make a reflection measurement on PORT 1 (in the forward direction, S
instrument default setting.
❏ For ES analyzers, if you want to make a reflection measurement on PORT 2 (in the reverse direction,
S
), press:
22
Meas
Refl: REV S22 (B/R)
3. Set any other measurement parameters that you want for the device measurement: power, sweep type,
number of points, or IF bandwidth.
4. To access the measurement error-correction menus, press:
Cal
CALIBRATE MENU
5. If your calibration kit is different than the kit specified under the
Cal
CAL KIT
SELECT CAL KIT
If your type of calibration kit is not listed in the displayed menu, refer to
page
7-57.
6. To select a response correction, press:
Cal
CALIBRATE MENU
Connect the short or open calibration standard to the port you selected for the test port (PORT 1 for S
or PORT 2 for S
on ES analyzers).
22
NOTE
Include any adapters or cables that you will have in the device measurement. That is,
connect the standard device to the particular connector where you will connect your DUT.
6-12
(select your type of kit)
RESPONSE
), leave the
11
softkey, press:
Cal
CAL KIT [ ]
RETURN
"Modifying Calibration Kits" on
11