Check 8, Cpu Fuse High Lead Voltage; Check 9, Fuse And Line Switch - Agilent Technologies 4349B Service Manual

4-channel high resistance meter
Hide thumbs Also See for 4349B:
Table of Contents

Advertisement

Check 8, CPU Fuse High Lead Voltage

1. Turn on the 4349B.
2. Using a multimeter, check if the dc voltage at the high lead of the fuse F1 on the CPU Board
shown in Figure 4-10 is +5
3. If the voltage is in the range, replace the A2 CPU Board Assembly, otherwise replace the A1
Main Board Assembly.
If the multimeter reading is correct, fuse F1 on the A2 CPU Board Assembly is
Note
most suspicious. Replacing only the fuse may x the problem.
When replacing the A2 CPU Board Assembly, refer to \Replacing the A2 CPU
Note
Board Assembly" in Chapter 3 for handling the ROM and EEPROM on the A2
Assembly.

Check 9, Fuse and Line Switch

1. Check that the fuse on the 4349B rear panel is not blown.
2. Check that the line switch on the A5 AC Inlet Board Assembly is not defective.
3. If passed, proceed to \Check 10, DC-DC Converter Input Voltage", otherwise replace the
defective line switch or blown fuse.
4-14
T roubleshooting
6
0.2 V.
Figure 4-10. Location of High Lead of A2F1 Fuse