Unit:temperature {C|F|K}Unit:temperature - Keysight Truevolt Series Operating And Service Manual

Digital multimeters
Table of Contents

Advertisement

UNIT:TEMPerature {C|F|K}
UNIT:TEMPerature?
Selects the units (°C, °F or Kelvin) to be used for all temperature measurements except the
SYSTem:TEMPerature?
query.
Parameter
{C|F|K}. Default: C.
Make a four-wire RTD measurement and return the result in °F:
UNIT:TEMP F
MEAS:TEMP? FRTD
Typical Response: +6.82320000E+01
The command also accepts CEL or FAR, but the query returns C or F.
l
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
See Also
CONFigure:TEMPerature
MEASure:TEMPerature?
[SENSe:]FUNCtion[:ON]
Keysight True volt Series Operating and Service Guide
Typical Return
C, F, or K
SCPI Programming Reference