SCPI Programming Reference
[SENSe:]{RESistance|FRESistance}:OCOMpensated {OFF|ON}
[SENSe:]{RESistance|FRESistance}:OCOMpensated?
Enables or disables offset compensation. Offset compensation removes the effects of small dc voltages in
the circuit being measured. The technique involves taking the difference between two resistance meas-
urements, one with the current source set to the normal value, and one with the current source set to a
lower value. Enabling Offset Compensation approximately doubles the reading time.
This parameter is common to 2-wire and 4-wire resistance measurements. Using the FRESistance version
of this command or query is identical to using the RESistance version.
Parameter
{ON|1|OFF|0}. Default: OFF.
Configure low-power 2-wire resistance measurements. Enable offset compensation. Make and read two
measurements:
RES:POW:LIM ON
RES:OCOM ON
SAMP:COUN 2
READ?
Typical Response: +4.05451008E-03,+4.97391062E-03
Applies only to the 34465A and 34470A.
l
Affects both 2-wire and 4-wire resistance measurements.
l
Applies only to resistance measurements on the 100Ω, 1 kΩ, 10 kΩ and 100 kΩ ranges.
l
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
352
Typical Return
0 (OFF) or 1 (ON)
Keysight True volt Series Operating and Service Guide