0 (Off) Or 1 (On); Typical Response: +1.04530000E+00,+1.04570000E+00 - Keysight Truevolt Series Operating And Service Manual

Digital multimeters
Table of Contents

Advertisement

[SENSe:]TEMPerature:NULL[:STATe] {ON|1|OFF|0}
[SENSe:]TEMPerature:NULL[:STATe]?
Enables or disables the null function for temperature measurements.
Parameter
{ON|1|OFF|0}. Default: OFF.
Configure 4-wire RTD measurements, using the null function to subtract 25°. Make and read two meas-
urements:
CONF:TEMP RTD
TEMP:NULL:STAT ON
VAL 25;SAMP:COUN 2
READ?

Typical Response: +1.04530000E+00,+1.04570000E+00

Enabling the scaling function also enables automatic null value selection
l
ature:NULL:VALue:AUTO
To set a fixed null value, use: [SENSe:]TEMPerature:NULL:VALue.
l
The instrument disables the null function after a Factory Reset (*RST), Instrument Preset
l
(SYSTem:PRESet), or CONFigure function.
[SENSe:]TEMPerature:NULL:VALue {< value >|MIN|MAX|DEF}
[SENSe:]TEMPerature:NULL:VALue? [{MIN|MAX|DEF}]
Stores a null value for temperature measurements.
Parameter
-1.0E15 to +1.0E15. Default: 0.
Configure 4-wire RTD measurements, using the null function to subtract 25°. Make and read two meas-
urements:
CONF:TEMP RTD
TEMP:NULL:STAT ON
VAL 25;SAMP:COUN 2
READ?
Typical Response: +1.04530000E+00,+1.04570000E+00
Specifying a null value disables automatic null value selection
l
ature:NULL:VALue:AUTO
To use the null value, the null state must be on
l
This parameter is set to its default value after a Factory Reset (*RST), Instrument Preset
l
(SYSTem:PRESet), or CONFigure function.
Keysight True volt Series Operating and Service Guide
0 (OFF) or 1 (ON)
ON).
+2.50000000E+01
OFF).
([SENSe:]TEMPerature:NULL:STATe
SCPI Programming Reference
Typical Return
([SENSe:]TEMPer-
Typical Return
([SENSe:]TEMPer-
ON).
361