Features and Functions
The Display and Digit Mask softkeys behave as they do in the
The Scale softkey specifies the horizontal scale:
Default sets the scale to equal the measurement range.
l
Manual allows you to configure the scale either as High and Low values or as a Span around a Center
l
value. For example, a scale that goes from a Low of -500 Ω to a High of 1000 Ω could also be specified
as a Center of 250 Ω with a Span of 1500 Ω.
Selecting a Secondary Measurement
Press 2nd Meas to select and display a secondary measurement. For example, for the DCV measurement
function, you can select ACV, Peak, or Pre-Math as the secondary measurement function. With ACV selec-
ted as secondary, the display shows the DCV measurement as a number near to top of the display, DCV in
the bar meter, and the ACV measurement above the bar meter: