Keysight Truevolt Series Operating And Service Manual page 512

Digital multimeters
Table of Contents

Advertisement

Input
Voltage Function
100 mV
-100 mV
1 V
-1 V
4 V
DC Volts
10 V
-10 V
100 V
-100 V
1000 V
-500 V
[1] Select 10 PLC.
Q
: Quick performance verification test point.
Input
Resistance Function
100 Ω
1 kΩ
10 kΩ
4-wire Ohms
100 kΩ
1 MΩ
10 MΩ
10 MΩ
2-wire Ohms
100 MΩ
[1] Select 10 PLC.
Q
: Quick performance verification test point.
Keysight True volt Series Operating and Service Guide
Values for 34465A DMM
Quick
[1]
Range
24 hours 90 days
Check
100 mV
± 6 µV
1 V
± 19 µV
± 70 µV
10 V
Q
± 130 µV
100 V
± 2.6 mV
Q
± 36 mV
1000 V
± 16 mV ± 23.5 mV ± 26 mV ± 33.5 mV
Values for 34465A DMM
Quick
[1]
Range
24 hours 90 days
Check
100 Ω
1 kΩ
Q
10 kΩ
± 250 mΩ ± 350 mΩ ± 450 mΩ ± 550 mΩ
100 kΩ
1 MΩ
10 MΩ
10 MΩ
Q
100 MΩ
Error from Nominal
1 year
± 7.5 µV
± 8.5 µV
± 29 µV
± 39 µV
± 120 µV ± 160 µV ± 220 µV
± 240 µV ± 340 µV ± 490 µV
± 4.1 mV ± 4.6 mV ± 6.1 mV
± 51 mV
± 56 mV
Error from Nominal
1 year
± 6 mΩ
± 9 mΩ
± 10 mΩ
± 25 mΩ
± 35 mΩ
± 45 mΩ
± 2.5 Ω
± 3.5 Ω
± 4.5 Ω
± 25 Ω
± 65 Ω
± 75 Ω
± 1.1 kΩ
± 2.1 kΩ
± 2.6 kΩ
± 101 kΩ ± 201 kΩ ± 301 kΩ ± 401 kΩ
Performance Verification
2 years
± 10 µV
± 54 µV
± 71 mV
2 years
± 11 mΩ
± 55 mΩ
± 5.5 Ω
± 85 Ω
± 3.1 kΩ
513

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

34461a34465a34470a34460a

Table of Contents