Keysight Truevolt Series Operating And Service Manual page 367

Digital multimeters
Table of Contents

Advertisement

SCPI Programming Reference
Applies only to the 34465A and 34470A.
l
For this command, you must always specify the temperature in degrees Celsius regardless of the tem-
l
perature units currently selected (see
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
[SENSe:]TEMPerature:TRANsducer:TCouple:RJUNction:OFFSet:ADJust
{< temperature >|MIN|MAX|DEF}
[SENSe:]TEMPerature:TRANsducer:TCouple:RJUNction:OFFSet:ADJust? [{MIN|MAX|DEF}]
Allows you to make minor temperature adjustments to correct for the differences between the DMM
internal temperature measurement of the front connection and the actual temperature of the meas-
urement terminals.
Parameter
-20 °C to +20 °C. Default: 0 °C.
Configure thermocouple temperature measurements using the internal reference junction. Measure the
temperature inside the instrument. Assuming the temperature outside the instrument is 5 °C cooler than
the inside measurement, adjust the temperature by -5 °C. Then make and read one measurement:
CONF:TEMP TC,K
TEMP:TRAN:TC:RJUN:TYPE INT
TEMP:TRAN:TC:CHEC ON
SYST:TEMP?
TEMP:TRAN:TC:RJUN:OFFS:ADJ -5
READ?
Typical Response: +3.31785468E+00
Applies only to the 34465A and 34470A.
l
Important: Since the internal reference temperature is the temperature of the front connections, use
l
of the rear connections with the selection of the internal reference junction will have an unknown
error with no specified performance and is not recommended.
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
368
UNIT:TEMPerature
command).
+2.00000000E+01
Keysight True volt Series Operating and Service Guide
Typical Return

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

34461a34465a34470a34460a

Table of Contents