Installing the Optional GPIB Interface

This procedure should be performed by qualified service personnel only. Turn off the
power and remove all measurement leads and other cables, including the power cord,
from the instrument before continuing.
Tools Required
The following tools are required.
T10 Torx driver
l
Installation Procedure
Turn off the power and remove all measurement leads and other cables, including the power cord,
from the instrument before continuing.
Use the Torx driver to remove the screw from the GPIB cover plate. Retain the screw for use later in
this procedure. Then remove the cover plate by sliding it to the left.
Retain GPIB Cover Plate
After installing the GPIB option, retain the cover plate for use in the event that you ever
remove the GPIB option. The instrument should never be connected to power or inputs on
the measurement terminals without either the GPIB module or the cover plate securely cov-
ering the rear-panel opening.
Locate the connector, just inside the unit, for the cable attached to the GPIB module
Keysight True volt Series Operating and Service Guide
Service and Repair