Calibration Adjustment Process

The following general procedure is recommended to complete a full calibration adjustment.
1. Read
Test
Considerations.
2. Perform the
performance verification tests
3. Unsecure the instrument for calibration (see
4. Perform the calibration procedures (see
5. Secure the instrument against calibration (see
6. Note the security code and calibration count in the instrument's maintenance records.
7. Perform the
performance verification tests
Keysight True volt Series Operating and Service Guide
to characterize the instrument (incoming data).
Calibration
Security).
Calibration
Procedures).
Calibration
Security).
to verify the calibration.
Calibration Adjustment Procedures