CONFigure Subsystem

The CONFigure commands are the most concise way to configure measurements. Like the
ies, these commands use default measurement configuration values. However, these commands do not
automatically start measurements, so you can modify measurement attributes before initiating the meas-
urement.
Use
INITiate
Command Summary
CONFigure?
CONFigure:CAPacitance
CONFigure:CONTinuity
CONFigure:CURRent:{AC|DC}
CONFigure:DIODe
CONFigure:{FREQuency|PERiod}
CONFigure:{RESistance|FRESistance}
CONFigure:TEMPerature
CONFigure[:VOLTage]:{AC|DC}
CONFigure[:VOLTage][:DC]:RATio
Default Settings for the CONFigure Command
The CONFigure commands select the function, range and resolution in one command. Specify
< resolution > in the measurement's units (V, A, Hz, Ω, and so on). All other parameters are set to their
default values (below).
Measurement Parameter
AC Input Filter (bandwidth)
Autozero
Range
Samples per Trigger
Trigger Count
Trigger Delay
Trigger Source
Trigger Slope
Math Functions
Per-function Null State
Keysight True volt Series Operating and Service Guide
or
READ?
to initiate measurements.
Default Setting
20 Hz (medium filter)
OFF if resolution setting results in NPLC < 1
ON if resolution setting results in NPLC ≥ 1
AUTO (including voltage range for frequency and period measurements)
1 sample
1 trigger
Automatic delay
Immediate
NEGative
Disabled. Other parameters are unchanged.
Disabled
SCPI Programming Reference
MEASure?
quer-
243