CALibration Subsystem

Read the
reliability.
CALibration:ADC?
CALibration[:ALL]?
CALibration:COUNt?
CALibration:DATE?
CALibration:SECure:CODE
CALibration:SECure:STATe
CALibration:STORe
CALibration:STRing
CALibration:TEMPerature?
CALibration:TIME?
CALibration:VALue
CALibration:ADC?
Performs a low-level calibration of the ADC (analog-to-digital converter) circuitry.
You must unsecure the instrument to calibrate it or to change the calibration code or security code.
Parameter
Typical Return
(none)
0 (successful) or 1 (unsuccessful)
Calibrate the ADC:
CAL:ADC?
This is the first step in a re-calibration sequence, and must be done before doing the offset calibration.
l
Keysight True volt Series Operating and Service Guide
calibration procedure
before calibrating. Improper calibration can reduce accuracy and
SCPI Programming Reference
237