MEASure[:VOLTage]:{AC|DC}? [{< range >|AUTO|MIN|MAX|DEF} [, {< resolution >|MIN|MAX|DEF}]]
Sets all measurement parameters and trigger parameters to their
urements and immediately triggers a measurement. The results are sent directly to the instrument's out-
put buffer.
The maximum range parameter (MAX) is 1000 V. However, the SAFETY LIMIT on the front
and rear HI/LO input terminals is 750 VAC (rms). The rms voltage is waveform dependent.
A sine wave is limited to 750 VAC (rms), but a 1000 Vpk square wave is safe. Connection
to AC mains power is further limited to CAT II (300 V). See
Information
instrument.
<range>: 100 mV, 1 V, 10 V, 100 V, 1000 V, AUTO (default) or DEFault
<resolution> (AC): optional and ignored; fixed at 6½ digits.
Resolution Table
<resolution> (DC): See
NPLC.
The default is equivalent to 10 PLC.
Specify <resolution> in the measurement's units (V, A, Hz, Ω, and so on).
Configure DC voltage measurements using the 10 V range with 1 mV resolution.Then make and read one
measurement:
MEAS:VOLT:DC? 10,0.001
Typical Response:+8.54530000E+01
You can let autoranging select the measurement range, or you can manually select a fixed range.
l
Autoranging conveniently selects the range for each measurement based on the input signal. For fast-
est measurements, use manual ranging (autoranging may require additional time for range selection).
Autoranging (AUTO or DEFault), will generate an error if you specify a < resolution > because the instru-
l
ment cannot accurately resolve the integration time (especially if the input continuously changes). If
your application requires autoranging, specify DEFault for the < resolution > or omit the < resolution >
altogether.
Autoranging goes down a range at less than 10% of range and up a range at greater than 120% of
l
range.
If the input signal is greater than can be measured on the specified manual range, the instrument dis-
l
plays the word Overload on front panel and returns "9.9E37" from the remote interface.
Keysight True volt Series Operating and Service Guide
for more information on the safety features and the safe operation of this
Parameter
Range, Resolution and
or
SCPI Programming Reference
default values
for AC or DC voltage meas-
Safety and Regulatory
Typical Return
(none)
289

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

34461a34465a34470a34460a

Table of Contents