SCPI Programming Reference
TRIGger:DELay:AUTO {ON|1|OFF|0}
TRIGger:DELay:AUTO?
Disables or enables automatic trigger delay. If enabled, the instrument determines the delay based on func-
tion, range, and integration time or bandwidth.
Parameter
{ON|1|OFF|0}. Default: ON.
Return 5 DC voltage measurements, with an automatic delay between each measurement.
CONF:VOLT:DC 10
SAMP:COUN 5
TRIG:DEL 2
READ?
Typical Response:
+4.27230000E+00,+4.27150000E+00,+4.27190000E+00,+4.27170000E+00,+4.27200000E+00
Selecting a specific trigger delay using
l
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
TRIGger:LEVel {< level >|MIN|MAX|DEF}
TRIGger:LEVel? [{MIN|MAX|DEF}]
Sets the level on which a trigger occurs when level triggering is enabled
INTernal).
It is recommended that all triggered measurements be made using an appropriate fixed
manual range. That is, turn autorange off
a fixed range using the SENSe:<function>:RANGe, CONFigure, or
Parameter
<level> (see bullet points below). Default: 0.
Configures dc voltage measurements, 1 V range, 0.1 mV resolution. Select internal trigger source. Set the
trigger level to 0.75 volts, positive slope. INIT puts the instrument in the "wait for trigger" mode.
CONF:VOLT:DC 1,0.0001
TRIG:SOUR INT
TRIG:LEV 0.75
TRIG:SLOP POS
INIT
Applies only to the 34465A and 34470A, and requires the DIG option.
l
Supported only for ac and dc voltage, ac and dc current, frequency, period, 2-wire resistance, 4-wire
l
resistance, RTD and thermistor temperature measurements. Resistance and temperature meas-
urements must have offset compensation and low power disabled.
To use the trigger level you must select INTernal as the trigger source with the
l
mand.
430
0 (OFF) or 1 (ON)
TRIGger:DELay
disables the automatic trigger delay.
(SENSe:<function>:RANGe:AUTO
+2.00000000E+01
Keysight True volt Series Operating and Service Guide
Typical Return
(TRIGger:SOURce
set to
OFF), or set
MEASure
command.
Typical Return
TRIGger:SOURce
com-