SCPI Programming Reference
CONFigure:{RESistance|FRESistance} [{< range >|AUTO|MIN|MAX|DEF} [, {< resolution >|MIN|MAX|DEF}]]
Sets all measurement parameters and trigger parameters to their
or 2-wire (RESistance) resistance measurements. Also specifies the range and resolution.
<range>: 100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1 MΩ, 10 MΩ, 100 MΩ, 1
GΩ, AUTO,or DEFault. Default: AUTO.
Resolution Table
<resolution>: See
NPLC.
The default is equivalent to 10 PLC.
Specify <resolution> in the measurement's units (V, A, Hz, Ω, and
so on).
Configure 4-wire resistance measurements using the 100 Ω range with default resolution. Make and read
two measurements:
CONF:FRES 100
SAMP:COUN 2
READ?
Typical Response: +8.54530000E+01,+8.54520000E+01
The 1 GΩ range applies only to the 34465A and 34470A.
l
You can let autoranging select the measurement range, or you can manually select a fixed range.
l
Autoranging conveniently selects the range for each measurement based on the input signal. For fast-
est measurements, use manual ranging (autoranging may require additional time for range selection).
Autoranging (AUTO or DEFault), will generate an error if you specify a < resolution > because the instru-
l
ment cannot accurately resolve the integration time (especially if the input continuously changes). If
your application requires autoranging, specify DEFault for the < resolution > or omit the < resolution >
altogether.
Autoranging goes down a range at less than 10% of range and up a range at greater than 120% of
l
range.
If the input signal is greater than can be measured on the specified manual range, the instrument dis-
l
plays the word Overload on front panel and returns "9.9E37" from the remote interface.
Use
READ?
or
INITiate
l
See Also
CONFigure?
MEASure:{RESistance|FRESistance}?
250
Parameter
Range, Resolution and
or
to start the measurement.
default values
for 4-wire (FRESistance)
Typical Return
+8.54530000E+01