SCPI Programming Reference
CALCulate:AVERage:CLEar[:IMMediate]
Clears all computed statistics: minimum, maximum, average, peak-to-peak, count and standard devi-
ation.
Parameter
Typical Return
(none)
(none)
Clear the stored statistical data:
CALC:AVER:CLE
Does not clear measurements in reading memory.
l
Statistics are cleared when the measurement function changes or when any of these commands is
l
executed:
CALCulate:AVERage:STATe ON
CALCulate:AVERage:CLEar
INITiate
MEASure:< function >?
READ?
*RST
SYSTem:PRESet
To clear statistics, limits, histogram data, and measurement data, use CALCulate:CLEar[:IMMediate].
l
CALCulate:AVERage[:STATe] {ON|1|OFF|0}
CALCulate:AVERage[:STATe]?
Enables or disables statistics computation.
Parameter
{ON|1|OFF|0}. Default: OFF.
Return the statistics of 100 frequency measurements:
CONF:FREQ 1.0E3,.001
SAMP:COUN 100
CALC:AVER:STAT ON
INIT
*WAI
CALC:AVER:ALL?
Typical Response: +1.00520000E+03,+1.00512000E+03,+1.00527000E+03,+4.13500000E-01
232
Typical Return
0 (OFF) or 1 (ON)
Keysight True volt Series Operating and Service Guide