Keysight Truevolt Series Operating And Service Manual page 475

Digital multimeters
Table of Contents

Advertisement

SCPI Programming Reference
Data Settings
Binary Block Data Format (34465A/70A only):
Border
Data
Data Transfer (34465A/70A only):
Separator
Header
Reading Limits
The Error Queue is cleared at power on. It is not cleared by *RST,
SYSTem:PRESet, or a front panel Preset.
Input/Output Configuration
1
Enable Interfaces:
• LAN
• GPIB
• USB
2
LAN Settings:
• DHCP
• Auto IP
• IP Address
• Subnet Mask
• Default Gateway
• DNS Server
• Host Name
1
• LAN Services
GPIB Settings:
• GPIB Address
1 Interface enable or LAN service changes take effect only after you cycle power.
2 LAN setting changes require a LAN restart. From SCPI you must cycle power.
3 Where nnnnn represents the last 5 digits of the instrument serial number.
The instrument uses LAN port 5024 for SCPI Telnet sessions and port
5025 for SCPI Socket sessions.
476
Factory Setting
Normal
ASCII, 9
Comma
Off
Off
Factory Setting
• Enabled
• Enabled
• Enabled
• On
• On
• 169.254.4.61
• 255.255.0.0
• 0.0.0.0
• 0.0.0.0
3
• K-34460A-nnnnn (for 34460A)
• K-34461A-nnnnn (for 34461A)
• K-34465A-nnnnn (for 34465A)
• K-34470A-nnnnn (for 34470A)
• Enable All
(VISA LAN, Sockets, Telnet, Web Server, mDNS, HiSLIP)
• 22
Keysight True volt Series Operating and Service Guide