SCPI Programming Reference

MEASure Subsystem

The MEASure queries are the easiest way to program measurements because they always use default meas-
urement parameters. You set the function, range, and resolution in one command, but you cannot change
other parameters from their default values. The results are sent directly to the instrument's output buffer.
A MEASure query is functionally equivalent to sending CONFigure followed immediately by READ?. The
difference is that CONFigure commands allow you to change parameters between the CONFigure and the
READ?
Command Summary
MEASure:CAPacitance?
MEASure:CONTinuity?
MEASure:CURRent:{AC|DC}?
MEASure:DIODe?
MEASure:{FREQuency|PERiod}?
MEASure:{RESistance|FRESistance}?
MEASure:TEMPerature?
MEASure[:VOLTage]:{AC|DC}?
MEASure[:VOLTage][:DC]:RATio?
Default Settings for MEASure?
With the MEASure? queries, you can select the function, range and resolution in one command. All other
parameters are set to their default values (below).
Measurement Parameter
AC Input Filter (bandwidth)
Autozero
Range
Samples per Trigger
Trigger Count
Trigger Delay
Trigger Source
Trigger Slope
Math Functions
Per-function Null State
280
Default Setting
20 Hz (medium filter)
OFF if resolution setting results in NPLC < 1
ON if resolution setting results in NPLC ≥ 1
AUTO (including voltage range for frequency and period measurements)
1 sample
1 trigger
Automatic delay
Immediate
NEGative
Disabled. Other parameters are unchanged.
Disabled
Keysight True volt Series Operating and Service Guide