SYSTem:ACALibration:DATE?
Returns the date of the last autocalibration in the format yyyy,mm,dd .
Parameter
Typical Return
(none)
2014,4,26
Return the autocalibration date:
SYST:ACAL:DATE?
(34465A/70A only)
l
You can perform this query regardless of whether the instrument is secured.
l
The date is based on the instrument's real-time clock. Set the date for the instrument's real-time clock
l
with SYSTem:DATE.
The real-time clock does not adjust itself for time zone changes or daylight savings time. Be sure to set
l
the date and time when you first receive your instrument.
This setting is non-volatile; it will not be changed by power cycling or
l
Use
*CAL?
to perform autocal.
l
See Also
SYSTem:DATE
SYSTem:ACALibration:TEMPerature?
Returns the temperature of the last autocalibration in °C.
Parameter
Typical Return
(none)
+2.42850208E+001
Return the autocalibration temperature:
SYST:ACAL:TEMP?
(34465A/70A only)
l
You can perform this query regardless of whether the instrument is secured.
l
This setting is non-volatile; it will not be changed by power cycling or
l
Use
*CAL?
to perform autocal.