DC Voltage Gain Calibration

This calibration procedure applies to the 34465A and 34470A only.
Review
Test Considerations
Configuration: DC Voltage
1. Configure each function and range in the order shown in the table below.
2. Apply the input signal shown in the Input column.
3. Enter the actual applied input voltage amplitude (see
Succeeded message indicates success; if the display shows Calibration Step Failed, check the input
value, range, function, and entered calibration value and repeat the calibration step.
4. Repeat steps 1 through 3 for each gain calibration point shown in the table.
5. Store the new calibration constants (see
Each range takes less than 4 seconds to complete.
Instrument Settings
Input Function
10 V
-10 V
0.1 V
DCV
-0.1 V
1 V
The 100V and 1000V ranges are adjusted later in the
ibration.
Keysight True volt Series Operating and Service Guide
and
Gain Calibration Considerations
Storing the Calibration
Range
10 V
10 V
100 mV
100 mV
1 V
Calibration Adjustment Procedures
before beginning this procedure.
Entering Calibration
Values). A Calibration Step
Constants).
DC Voltage High Voltage Gain Cal-
Next Procedure
563