Performance Verification
Input
Current Function
2 A
1 A
100 mA
DC Current
10 mA
1 mA
[1] Select 10 PLC.
Q
: Quick performance verification test point.
Input
Voltage Function
100 mV
-100 mV
1 V
-1 V
4 V
DC Volts
10 V
-10 V
100 V
-100 V
1000 V
-500 V
[1] Select 10 PLC.
Q
: Quick performance verification test point.
Input
Resistance Function
100 Ω
1 kΩ
10 kΩ
4-wire Ohms
100 kΩ
1 MΩ
10 MΩ
10 MΩ
2-wire Ohms
100 MΩ
[1] Select 10 PLC.
Q
: Quick performance verification test point.
514
Values for 34470A DMM
Quick
[1]
Range
24 hours 90 days
Check
3 A
± 4.2 mA
1 A
Q
± 560 µA ± 800 µA ± 900 µA ± 1.1 mA
100 mA
± 14 µA
10 mA
Q
± 2.7 µA
1 mA
± 0.1 µA ± 0.35 µA ± 0.55 µA ± 0.65 µA
Values for 34470A DMM
Quick
[1]
Range
24 hours 90 days
Check
100 mV
± 6 µV
1 V
± 14 µV
± 52 µV
10 V
Q
± 100 µV ± 150 µV ± 180 µV ± 220 µV
100 V
± 2.6 mV ± 3.8 mV ± 4.4 mV ± 4.6 mV
Q
± 36 mV
1000 V
± 16 mV
Values for 34470A DMM
Quick
[1]
Range
24 hours 90 days
Check
100 Ω
1 kΩ
Q
10 kΩ
± 250 mΩ ± 350 mΩ ± 450 mΩ ± 550 mΩ
100 kΩ
1 MΩ
10 MΩ
10 MΩ
Q
100 MΩ
Error from Nominal
1 year
± 4.6 mA
± 35 µA
± 55 µA
± 5 µA
± 7 µA
Error from Nominal
1 year
± 7.5 µV
± 19 µV
± 24 µV
± 72 µV
± 84 µV ± 100 µV
± 48 mV ± 54 mV ± 56 mV
± 22 mV ± 25 mV ± 36 mV
Error from Nominal
1 year
± 6 mΩ
± 9 mΩ
± 10 mΩ
± 25 mΩ
± 35 mΩ
± 45 mΩ
± 2.5 Ω
± 3.5 Ω
± 4.5 Ω
± 25 Ω
± 65 Ω
± 75 Ω
± 1.1 kΩ
± 2.1 kΩ
± 2.6 kΩ
± 101 kΩ ± 201 kΩ ± 301 kΩ ± 401 kΩ
Keysight True volt Series Operating and Service Guide
2 years
± 5.2 mA
± 65 µA
± 8 µA
2 years
± 8 µV
± 29 µV
2 years
± 11 mΩ
± 55 mΩ
± 5.5 Ω
± 85 Ω
± 3.1 kΩ