Finishing Calibration - Keysight Truevolt Series Operating And Service Manual

Digital multimeters
Table of Contents

Advertisement

Calibration Adjustment Procedures

Finishing Calibration

To finish calibrating your instrument:
1. Remove all shorting blocks and connections from the instrument.
2. Reset the
calibration
3.
Secure the instrument
4. Record the
new calibration
5. Perform the
performance verification tests
554
message.
with the calibration password.
count.
to verify the calibration.
Keysight True volt Series Operating and Service Guide