Keysight Truevolt Series Operating And Service Manual page 325

Digital multimeters
Table of Contents

Advertisement

SCPI Programming Reference
[SENSe:]CURRent:{AC|DC}:TERMinals {3|10}
[SENSe:]CURRent:{AC|DC}:TERMinals?
Configures the AC or DC current measurement to measure the source on the 3 A or 10 A terminals. Note
that the 10 A terminal is not available on the 34460A.
Parameter
{3|10}. Default: 3.
Configure AC current measurements to use the 10 A terminal. Make and read two measurements:
CONF:CURR:AC
CURR:AC:TERM 10
SAMP:COUN 2
READ?
Typical Response: +7.81929394E+00,7.82013671E+00
Selecting the 10 A terminal with
l
rent measurements to use the 10 A range, but it does not affect the values of
{AC|DC}:RANGe
or [SENSe:]CURRent:{AC|DC}:RANGe:AUTO.
Changing the value of this parameter always changes the measurement range, which results in a
l
change of measurement resolution. The number of PLCs (NPLC) stays fixed, and this results in a
change of resolution in terms of the units of the measurement. For example, if you have a meas-
urement with 1 µA resolution on the 1 A range, changing to the 10 A terminals results in a resolution
of 10 µA.
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
326
+3 or +10
[SENSe:]CURRent:{AC|DC}:TERMinals
Keysight True volt Series Operating and Service Guide
Typical Return
forces the specified type of cur-
[SENSe:]CURRent:

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

34461a34465a34470a34460a

Table of Contents