CONFigure:{FREQuency|PERiod} [{< range >|MIN|MAX|DEF} [, {< resolution >|MIN|MAX|DEF}]]
Sets all measurement and trigger parameters to their
urements. Also specifies the frequency or period range and resolution.
Parameter
<range>: 3 Hz to 300 kHz. Default: 20 Hz. (FREQuency)
<range>: 3.33 µs to 333.33 ms. Default: 50 ms
(PERiod).
<resolution>: See table below.
Specify <resolution> in the measurement's units (V, A, Hz, Ω, and
so on).
Configure frequency measurements using the default aperture and the 200 Hz (fast) AC filter. Then make
and read one measurement:
CONF:FREQ
FREQ:RANG:LOW 200
READ?
Typical Response: +1.32130000E+03
The 1 ms aperture applies only to the 34465A and 34470A.
l
The frequency or period < range > parameter is required only to specify the measurement resolution
l
parameter. It is not necessary to send a new command for each new frequency or period to be meas-
ured. The < range > and < resolution > parameters affect the aperture (gate time) as follows:
Resolution (34460A/61A)
Not Applicable
100 ppm × <range> (MAXimum) 10 ppm × <range> (DEFault)
10 ppm × <range> (DEFault)
1 ppm × <range> (MINimum)
The input signal for frequency or period measurements has an AC voltage component. Use
l
{FREQuency|PERiod}:VOLTage:RANGe:AUTO
ure:{FREQuency|PERiod}
od}:VOLTage:RANGe
If the input voltage is too large for the selected voltage range (manual ranging), the instrument dis-
l
plays the word Overload on the front panel and returns "9.9E37" from the remote interface. Autor-
anging can be enabled for the input voltage.
Use
READ?
or
INITiate
l
See Also
CONFigure?
MEASure:{FREQuency|PERiod}?
Keysight True volt Series Operating and Service Guide
Resolution (34465A/70A)
100 ppm × <range> (MAXimum) 1 ms
1 ppm × <range>
0.1 ppm × <range> (MINimum) 1 s
to disable or enable voltage autoranging or use
to select voltage autoranging by default. Use
to select a fixed voltage range for frequency and period measurements.
to start the measurement.
SCPI Programming Reference
default values
for frequency or period meas-
Typical Return
(none)
[SENSe:]{FREQuency|PERi-
Aperture
10 ms
100 ms
[SENSe:]
CONFig-
249