CALibration:STRing "< string >"
CALibration:STRing?
Stores a message in calibration memory. Common messages include the last calibration date, calibration
due date, or calibration department contact information. You can perform this query regardless of
whether the instrument is secured.
You must unsecure the instrument to calibrate it or to change the calibration code or security code.
Quoted string up to 40 characters
May contain letters, numbers, spaces, and other common char-
acters.
CAL:STR "FOR CAL HELP, CALL JOE AT EXT 1234"
This string may be stored only from the remote interface, but you can read the message from the front
l
panel or remote interface.
Storing a calibration message overwrites the previous message.
l
This setting is non-volatile; it will not be changed by power cycling or
l
CALibration:TEMPerature?
Returns the temperature of the last calibration in °C.
Parameter
Typical Return
(none)
+2.42850208E+001
Return the calibration temperature:
CAL:TEMP?
You can perform this query regardless of whether the instrument is secured.
l
This setting is non-volatile; it will not be changed by power cycling or
l
CALibration:TIME?
Returns the amount of time since the last calibration in the format hh,mm,ss.sss .
Parameter Typical Return
(none)
20,15,30.000
Return the calibration time:
CAL:TIME?
You can perform this query regardless of whether the instrument is secured.
l
This setting is non-volatile; it will not be changed by power cycling or
l
Keysight True volt Series Operating and Service Guide
Parameter
SCPI Programming Reference
Typical Return
"FOR CAL HELP, CALL JOE AT EXT
1234"
(If no string stored, returns "")
*RST
or SYSTem:PRESet.
*RST
or SYSTem:PRESet.
*RST
or SYSTem:PRESet.