SCPI Programming Reference
CALibration:SECure:STATe {ON|1|OFF|0}[, < code >]
CALibration:SECure:STATe?
Unsecures or secures the instrument for calibration.
You must unsecure the instrument to calibrate it or to change the calibration code or security code.
Parameter
{ON|1|OFF|0}. Default: ON. 0 (OFF) or 1 (ON)
Unsecure calibration:
CAL:SEC:STAT OFF,MY_CODE_272
Secure calibration:
CAL:SEC:STAT ON
The < code > is optional to secure the instrument, but must be correct if provided.
l
When shipped from the factory, the instrument is secured, with the security code (calibration pass-
l
word) set to AT3446XA .
Front panel and remote interface calibration share the same code.
l
This setting is non-volatile; it will not be changed by power cycling or
l
CALibration:STORe
Takes calibration constants in volatile memory (CALibration:ALL?), and places them in nonvolatile
memory, where they are not changed by cycling power, *RST, or SYSTem:PRESet. Do this at the end of cal-
ibration to avoid losing changes.
You must unsecure the instrument to calibrate it or to change the calibration code or security code.
Parameter
(none)
Store calibration constants into non-volatile memory:
CAL:STOR
240
Typical Return
Typical Return
(none)
*RST
or SYSTem:PRESet.
Keysight True volt Series Operating and Service Guide