SCPI Programming Reference
[SENSe:]TEMPerature:TRANsducer:TCouple:TYPE {E|J|K|N|R|T}
[SENSe:]TEMPerature:TRANsducer:TCouple:TYPE?
Selects the thermocouple type for thermocouple measurements.
Parameter
{E|J|K|N|R|T} default J
Configure K-type thermocouple measurements. Select fixed reference junction and set reference junction
temperature to +20 °C. Enable thermocouple check. Make and read two measurements:
CONF:TEMP TC,K
TEMP:TRAN:TC:RJUN:TYPE FIX
TEMP:TRAN:TC:RJUN 20.0
TEMP:TRAN:TC:CHEC ON
SAMP:COUN 2
READ?
Typical Response: +3.31785468E+00,+3.61554515E+00
Applies only to the 34465A and 34470A.
l
Thermocouple measurements require a reference junction temperature (see
l
ature:TRANsducer:TCouple:RJUNction:TYPE
can use an internal measurement, or a known fixed junction temperature. By default, an internal ref-
erence junction temperature is used.
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
[SENSe:]TEMPerature:TRANsducer:{FTHermistor|THERmistor}:TYPE 5000
[SENSe:]TEMPerature:TRANsducer:{FTHermistor|THERmistor}:TYPE?
Selects the thermistor type for 2-wire and 4-wire temperature measurements. Choosing the THERmistor
or the FTHermistor command form makes no difference, as both commands affect the same parameter.
Parameter
5000 (only available choice)
Configure 4-wire thermistor measurements. Make and read two measurements:
CONF:TEMP FTH
TEMP:TRAN:FTH:TYPE 5000
SAMP:COUN 2
READ?
Typical Response: +1.04530000E+02,+1.04570000E+02
370
E, J, K, N, R, or T
command). For the reference junction temperature, you
Typical Return
+5000