Keysight Truevolt Series Operating And Service Manual page 411

Digital multimeters
Table of Contents

Advertisement

SCPI Programming Reference
SYSTem:COMMunicate:LAN:IPADdress "< address >"
SYSTem:COMMunicate:LAN:IPADdress? [{CURRent|STATic}]]
Assigns a static Internet Protocol (IP) address for the instrument. If DHCP is enabled (SYSTem:COMMu-
nicate:LAN:DHCP ON), the specified static IP address is not used. Contact your LAN administrator for
details.
If you change this setting, you must send
ate the new setting.
Parameter
nnn.nnn.nnn.nnn
Command: "
Query: {CURRent|STATic}. Default: CURRent.
Set a static IP address:
SYST:COMM:LAN:IPAD "169.254.149.35"
SYST:COMM:LAN:UPD
SYSTem:SECurity:IMMediate
l
On the 34460A, this requires option 34460A-LAN or option 3446LANU.
l
For the query form, specify "CURRent" (default) to read the value currently being used by the instru-
l
ment. Specify "STATic" to read the value currently stored in nonvolatile memory within the instrument
(may not be the actual address used by the instrument if DHCP is enabled).
This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instru-
l
ment Preset (SYSTem:PRESet).
This parameter is set to its default value when the instrument is shipped from the factory and after
l
SYSTem:SECurity:IMMediate.
SYSTem:COMMunicate:LAN:MAC?
Returns the instrument's Media Access Control (MAC) address as an ASCII string of 12 hexadecimal char-
acters (0-9 and A-F) enclosed in quotation marks.
Your LAN administrator may need the MAC address to assign a static IP address for this
device.
Parameter
Typical Return
(none)
"0030D3001041"
Return the MAC address:
SYST:COMM:LAN:MAC?
On the 34460A, this requires option 34460A-LAN or option 3446LANU.
l
The MAC address is also known as the link-layer address, the Ethernet (station) address, LANIC ID or
l
Hardware Address. This is an unchangeable 48-bit address assigned by the manufacturer to each
unique Internet device.
412
Typical Return
"169.254.149.35"
"
sets this parameter to its default value.
SYSTem:COMMunicate:LAN:UPDate
Keysight True volt Series Operating and Service Guide
to activ-

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

34461a34465a34470a34460a

Table of Contents