MEASure:CONTinuity?
Sets all measurement parameters and trigger parameters to their
immediately triggers a measurement. The results are sent directly to the instrument's output buffer.
Parameter
(none)
Configure the instrument for continuity measurements. Then make and read one measurement:
MEAS:CONT?
Typical Response: +1.32130000E-02
The range and resolution are fixed at 1 kΩ for continuity tests (a 2-wire resistance measurement).
l
The instrument beeps (if the beeper is enabled) for each measurement less than or equal to the con-
l
tinuity threshold (≤10 Ω), and the actual resistance measurement appears on the display.
From 10 Ω to 1.2 kΩ, the instrument displays the actual resistance measurement with no beep. Above
l
1.2 kΩ, the instrument displays "OPEN" with no beep.
The FETCh?, READ?, and
l
its value.