Frequency Accuracy Verification - Keysight Truevolt Series Operating And Service Manual

Digital multimeters
Table of Contents

Advertisement

Frequency Accuracy Verification

Configuration: Frequency (CONFigure:FREQuency DEF, MIN)
1. Make sure you have read
2. Select the Frequency function, default range, 3 Hz filter, and 1 second aperture.
3. Connect the Keysight 33500B Series Function/Arbitrary Waveform Generator to the input terminals.
For the 34461A/65A/70A, use the front panel input terminals and select the front input terminals with
the Front/Rear switch.
4. Select each range in the order shown below. Provide the indicated input voltage and frequency. Com-
pare measurement results to the test limits shown below. (Be certain to allow for appropriate source
settling.)
Input
Vrms
Frequency
0.1 Vrms
10 Hz
0.01 Vrms
300 kHz
Input
Vrms
Frequency
0.1 Vrms
10 Hz
0.01 Vrms
300 kHz
Input
Vrms
Frequency
0.1 Vrms
10 Hz
0.01 Vrms
300 kHz
Q:
Quick performance verification test point.
Keysight True volt Series Operating and Service Guide
Test
Considerations.
Values for 34460A DMM
Quick
Range
24 hours 90 days
Check
1 Vrms
0.1 Vrms
Q
± 60 Hz
Values for 34461A DMM
Quick
Range
24 hours 90 days
Check
1 Vrms
0.1 Vrms
Q
± 60 Hz
Values for 34465A and 34470A DMMs
Quick
Range
24 hours 90 days
Check
1 Vrms
0.1 Vrms
Q
± 60 Hz
Performance Verification
Error from Nominal
1 year
2 years
± 3 mHz
± 3.5 mHz
± 240 Hz ± 360 Hz ± 510 Hz
Error from Nominal
1 year
2 years
± 3 mHz
± 3.5 mHz
± 180 Hz ± 300 Hz ± 450 Hz
Error from Nominal
1 year
2 years
± 3 mHz
± 150 Hz ± 210 Hz ± 270 Hz
515

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

34461a34465a34470a34460a

Table of Contents