Keysight Truevolt Series Operating And Service Manual page 511

Digital multimeters
Table of Contents

Advertisement

Performance Verification
Input
Current Function
2 A
1 A
100 mA
DC Current
10 mA
1 mA
100 µA
[1] Select 10 PLC.
Q
: Quick performance verification test point.
Input
Resistance Function
100 Ω
1 kΩ
10 kΩ
4-wire Ohms
100 kΩ
1 MΩ
10 MΩ
100 MΩ
2-wire Ohms 100 MΩ
[1] Select 10 PLC.
Q
: Quick performance verification test point.
Input
Current Function
2 A
1 A
100 mA
DC Current
10 mA
1 mA
[1] Select 10 PLC.
Q
: Quick performance verification test point.
512
Values for 34461A DMM
Quick
[1]
Range
24 hours
Check
3 A
± 4.2 mA
1 A
Q
± 560 µA
100 mA
± 14 µA
10 mA
Q
± 2.7 µA
1 mA
± 0.13 µA ± 0.36 µA
100 µA
± 0.03 µA ± 0.065 µA ± 0.075 µA ± 0.085 µA
Values for 34461A DMM
Quick
[1]
Range
24 hours 90 days
Check
100 Ω
1 kΩ
Q
10 kΩ
± 250 mΩ ± 900 mΩ
100 kΩ
1 MΩ
10 MΩ
Q
Values for 34465A DMM
Quick
[1]
Range
24 hours 90 days
Check
3 A
± 4.2 mA ± 4.6 mA ± 4.6 mA ± 5.2 mA
1 A
Q
± 560 µA ± 800 µA ± 900 µA ± 1.1 mA
100 mA
± 14 µA
10 mA
Q
± 2.7 µA
1 mA
± 0.1 µA ± 0.35 µA ± 0.55 µA ± 0.65 µA
Error from Nominal
90 days
1 year
± 4.6 mA
± 4.6 mA
± 900 µA
± 1.1 mA
± 35 µA
± 55 µA
± 5 µA
± 7 µA
± 0.56 µA
Error from Nominal
1 year
± 6 mΩ
± 12 mΩ
± 14 mΩ
± 25 mΩ
± 90 mΩ ± 110 mΩ ± 130 mΩ
± 1.1 Ω
± 2.5 Ω
± 9 Ω
± 11 Ω
± 30 Ω
± 90 Ω
± 110 Ω
± 1.6 kΩ
± 2.1 kΩ
± 4.1 kΩ
± 310 kΩ
± 810 kΩ
Error from Nominal
1 year
± 35 µA
± 55 µA
± 5 µA
± 7 µA
Keysight True volt Series Operating and Service Guide
2 years
± 5.2 mA
± 1.3 mA
± 65 µA
± 8 µA
± 0.66 µA
2 years
± 16 mΩ
± 1.3 Ω
± 13 Ω
± 130 Ω
± 6.1 kΩ
2 years
± 65 µA
± 8 µA