If you are using a wide aperture (NPLC or aperture mode) and the signal level varies significantly within
l
the aperture, you may get an unexpected variation in the magnitude of the triggered measurement
from the specified trigger level.
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
Specify < level > in the fundamental units for the function (volts, ohms, amps, hertz, or seconds). The
l
table below shows the allowable values for < level > for each measurement function, range, and input
terminals (for DCI and ACI).
Measurement Func-
tion
DCV
DCI
RESistance/FRESistance All ranges, fixed range
ACV
ACI
FREQuency
PERiod
*N/A = Not Applicable.
See Also
TRIGger:DELay
TRIGger:SOURce
Keysight True volt Series Operating and Service Guide
Range/Input Ter-
minals
100 mV to 100 V
±120% of range
ranges, fixed range
1000 V range, fixed
±1000 V
range
Autorange
±1000 V
3 A terminals, fixed
±120% of range
range
3 A terminals, autor-
±3.6 A
ange
10 A terminals
±12 A
0 to +120% of range
Autorange
0 to +1.2 GΩ
100 mV to 100 V
0 to +120% of range
ranges, fixed range
750 V range, fixed
0 to +1000 V
range
Autorange
0 to +1000 V
3 A terminals, fixed
0 to +120% of range
range
3 A terminals, autor-
0 to +3.6 A
ange
10 A terminals
0 to +12 A
*N/A
3 Hz to 300 kHz
*N/A
3.33333333 µs to 0.333333333 s
SCPI Programming Reference
<level> Parameter Value
431