Features and Functions
Step 5: Press Gate Time and choose the measurement aperture (integration time) of 1 ms (34465A/70A
only) 10 ms, 100 ms (default), or 1 s.
Step 6: (34465A/70A only) Press Timeout to control how long the instrument waits before timing out on
a frequency or period measurement when no signal is present. When set to 1s, the instrument waits 1
second before timing out. When set to Auto, the wait time varies with AC filter bandwidth; for the faster
bandwidths, the instrument waits a shorter time before timing out and returning 0.0. This is advantageous
in manufacturing test systems where a DUT failure may result in no signal; in this case, the failure can be
detected sooner, speeding up test throughput.
For accurate displayed statistics of AC measurements in Front Panel mode, the default manual trigger delay
( [Acquire] > Delay Man ) must be used.