Automatic Trigger Delays - Keysight Truevolt Series Operating And Service Manual

Digital multimeters
Table of Contents

Advertisement

SCPI Programming Reference

Automatic Trigger Delays

If TRIGger:DELay:AUTO is on, the instrument selects the trigger delay for you, as shown in the tables
below.
All frequency and period measurements have an automatic trigger delay of 1 second.
The continuity and diode functions ignore the trigger delay.
The 2-wire and 4-wire temperature measurements use the corresponding 2-wire and 4-wire
resistance measurement delays.
DC Voltage Default Delays
0.001
PLC
Aperture (60 Hz Power)
20 µs
Aperture (50 Hz Power)
Range
All
1
34465A/70A with DIG option only.
2
34465A/70A only.
DC Current Default Delays
0.001
PLC
Aperture (60 Hz Power)
20 µs
Aperture (50 Hz Power)
Range
All
1
34465A/70A with DIG option only.
2
34465A/70A only.
456
1
1
1
0.002
0.006
0.06
0.02
1
1
1
40 µs
100 µs
0.3 ms 1 ms
Delay
100 µs
130 µs
1
1
1
0.002
0.006
0.06
0.02
1
1
1
40 µs
100 µs
0.3 ms 1 ms
Delay
1 ms
2
0.2
1
10
100
16.7 ms 0.167 s 1.67 s
2
3 ms
20 ms
0.2 s
2 s
160 µs
2
0.2
1
10
100
16.7 ms 0.167 s 1.67 s
2
3 ms
20 ms
0.2 s
2 s
1.5 ms
Keysight True volt Series Operating and Service Guide